2010 (Swedish)Independent thesis Basic level (professional degree), 10 credits / 15 HE credits
Student thesis
Abstract [sv]
International Society for Clinical Densitometry (ISCD) rekommenderar att precisionsbestämning av bentäthetsmätning görs på varje klinik för att bedöma reproducerbarheten. Bentäthetsmätning görs för att diagnosticera osteoporos, följa upp behandling och förutsäga frakturrisk. I studien användes Dual energy x-ray absorptiometry (DXA) för att utföra dubbla mätningar på redan inbokade patienter på klinisk fysiologi, Skånes Universitetssjukhus (SUS), Lund. 105 patienter ingick i studien. Mätningarna utfördes på helkropp, totalhöft, lårbenshals och ländrygg. Helkroppsmätningar ingick i en interindividuell studie och de övriga ingick i intraindividuella studier. Reproducerbarheten uttrycktes som minsta signifikanta skillnaden (LSC), root mean square standardavvikelse (RMS SD) och variationskoefficient (%CV). Reproducerbarheten i studien var god med låga värden för LSC och RMS SD. %CV understeg de av ISCD rekommenderade maxvärdena, vilka är 1,8 % för totalhöft, 2,5 % för lårbenshals och 1,9 % för ländrygg.
Abstract [en]
It is recommended by the International Society for Clinical Densitometry (ISCD) that every clinic performs a precision assessment of bone densitometry to evaluate the reproducibility. Bone densitometry is used for diagnosis of osteoporosis, to monitor response to treatment and to assess patients’ risk of fractures. Dual energy x-ray absorptiometry (DXA) was used to perform double scans of patients already booked for examination at the section of clinical physiology, Skåne University Hospital (SUS), Lund. 105 patients were included in the study. Measurements were made at whole body, total hip, femoral neck and lumbar spine. The whole body scans were included in an interindividual study, whereas the others were included in intraindividual studies. The reproducibility was expressed as least significant change (LSC), root mean square standard deviation (RMS SD) and coefficient of variation (%CV). The reproducibility in the study was good, with low values for LSC and RMS SD. %CV fell below the maximal values recommended by the ISCD, which are 1,8 % for total hip, 2,5 % for femoral neck and 1,9 % for lumbar spine.
Place, publisher, year, edition, pages
Malmö högskola/Hälsa och samhälle , 2010.
Keywords [sv]
bentäthetsmätning, dual energy x-ray absorptiometry, precisionsbestämning, osteoporos, reproducerbarhet, International Society for Clinical Densitometry
National Category
Medical and Health Sciences
Identifiers
URN: urn:nbn:se:mau:diva-24534Local ID: 11640OAI: oai:DiVA.org:mau-24534DiVA, id: diva2:1485954
Educational program
HS Biomedicinsk laboratorievetenskap
2020-11-022020-11-022022-06-27Bibliographically approved