Effect of Implant Design on Fracture Resistance, Mode, and Crystal Structure of One- and Two-Piece Zirconia Implants With Ceramic or Titanium Abutment Screws: An In Vitro Study
Malmö universitet, Odontologiska fakulteten (OD). Department of Oral and Maxillofacial Surgery, Faculty of Dental Medicine, Umm Al-Qura University, Makkah, Saudi Arabia.ORCID-id: 0000-0001-7726-7086
Malmö universitet, Odontologiska fakulteten (OD).ORCID-id: 0000-0001-6260-473X
The Nordic Institute of Dental Materials, NIOM, Oslo, Norway.ORCID-id: 0009-0002-8409-1634
The Nordic Institute of Dental Materials, NIOM, Oslo, Norway.ORCID-id: 0009-0005-2732-5467
Visa övriga samt affilieringar
2026 (Engelska)Ingår i: Clinical Oral Implants Research, ISSN 0905-7161, E-ISSN 1600-0501, Vol. 37, nr 9, s. 1081-1095Artikel i tidskrift (Refereegranskat) Published
Abstract [en]

Objectives: To evaluate the fracture resistance, fracture modes, and crystalline phase changes of zirconia one-piece implants (OPI) and bone-level (BLTPI) or tissue-level, two-piece implants (TLTPI), with ceramic (C) or titanium (T) abutment screws after dynamic preloading in a moist environment. Materials and Methods: Dynamic preloading under moist conditions and subsequent static load-to-fracture testing were performed to evaluate fracture resistance and fracture mode. Scanning electron microscopy (SEM) analysis of fracture surfaces was employed to supplement the observations of macroscopic fracture modes. Crystalline phase changes were assessed using micro-Raman spectroscopy. Specimen preparations were performed in accordance with ISO 14801:2016. Five different ceramic (zirconia) implant designs were divided into the following groups: OPI, BLTPI (BLC, BLT), TLTPI (TLC, TLT). One-way and two-way ANOVA tests were used for comparisons and factor effect analysis. Results: Fracture resistance (N) and bending moment (N cm) differed significantly among the implant groups (p < 0.001), following the order TLTPI (775.2 ± 71 N) > OPI (605.7 ± 37 N) > BLTPI (498.3 ± 93 N). Within the two-piece implants (TPI), tissue-level designs (TL) outperformed bone-level designs (BL), while designs with different abutment screw materials showed no significant difference (p = 0.982). Conclusions: Within the limitations of this in vitro study, the evaluated zirconia-based OPI, BLTPI, and TLTPI systems demonstrate sufficient fracture resistance to support further clinical investigation. Implant-level design (TL vs. BL) influences mechanical performance, whereas the influence of abutment screw material (ceramic vs. titanium) is less pronounced and varies according to the implant-level design.

Ort, förlag, år, upplaga, sidor
John Wiley and Sons Inc , 2026. Vol. 37, nr 9, s. 1081-1095
Nyckelord [en]
bone-level implants, dental implants, implant design, implant fracture, tissue-level implants, zirconia implants
Nationell ämneskategori
Odontologi
Identifikatorer
URN: urn:nbn:se:mau:diva-87151DOI: 10.1111/clr.70149ISI: 001792673700001PubMedID: 42288470Scopus ID: 2-s2.0-105041489976OAI: oai:DiVA.org:mau-87151DiVA, id: diva2:2087707
Tillgänglig från: 2026-07-22 Skapad: 2026-07-22 Senast uppdaterad: 2026-09-23Bibliografiskt granskad
Ingår i avhandling
1. Zirconia dental implants: the influence of design on clinical outcomes, stress distribution and fracture resistance
Öppna denna publikation i ny flik eller fönster >>Zirconia dental implants: the influence of design on clinical outcomes, stress distribution and fracture resistance
2026 (Engelska)Doktorsavhandling, sammanläggning (Övrigt vetenskapligt)
Abstract [en]

In recent years, zirconia dental implants have been attracting increasingly more recognition and interest. Commercially, these implants are available in either one- piece (OPI) or two-piece (TPI) designs. The TPI design is categorised as either bone- level (BL) or tissue-level (TL) and is connected to the abutment with either a titanium or a ceramic abutment screw. Due to zirconia's brittle nature and high elastic modulus, the design of zirconia implants and their implant to abutment connection may influence load transfer from the prosthesis to the implant and surrounding bone, thereby affecting stress distribution, fracture resistance, and ultimately clinical survival.

Existing research on stress distribution, fracture resistance, and clinical outcomes of various zirconia implant designs and material combinations is limited. Therefore, this thesis aims to evaluate the clinical outcomes, stress concentrations, and fracture resistance of different designs of zirconia dental implants. In line with the aim, the thesis comprises three studies: Study I – A systematic review of clinical studies that evaluated the clinical outcomes of various zirconia implant designs. The findings showed that the one-piece implant had a promising 5-year clinical survival rate (95%), with no difference across implant diameters and lengths. The design with an acid-etched surface showed higher survival compared to other surfaces.

Study II – A finite element analysis that evaluated the location of high-stress levels and concentrations in various zirconia implant designs. The findings showed that OPI and TPI designs showed differing stress distributions. TPIs demonstrated lower stress within the implant body but higher stress transfer to the specimen holder. Among TPI designs, Implant level (TL vs BL) and abutment screw design/material influenced the stress distribution within the implant-abutment complex.

Study III – An in vitro laboratory study that evaluated the fracture resistance, fracture mode, and crystal phase change after dynamic preloading of various zirconia dental implant designs. The findings showed that the evaluated zirconia-based OPI, BLTPI, and TLTPI systems possessed adequate fracture resistance for further clinical evaluation. Fracture resistance and fracture mode were influenced by implant-level design (TL vs BL), whereas the abutment screw design/material (ceramic vs titanium) had a less pronounced influence and depended on the implant- level design.

Overall, within the limitations of this thesis, zirconia dental implants should not be regarded as a uniform category. Differences in implant design, including OPI and TPI, TL and BL configurations, and implant-abutment connection design, influence clinical outcomes, stress distribution, and fracture resistance. Clinical evidence is currently better established for OPI, which has shown promising clinical outcomes, while evidence for TPI remains limited. Mechanically, TL TPI shows the highest fracture resistance among the evaluated designs. However, further long-term clinical studies are needed to confirm these findings, particularly for TPI.

Abstract [sv]

På senare år har dentala implantat av zirkoniumdioxid fått ökad uppmärksamhet inom tandvården. Kommersiellt finns dessa implantat tillgängliga i olika designer, antingen som en del där implantat och distans utgör en enhet (one piece implant, OPI) eller i två delar (two piece implant, TPI) som bestående av implantat och distans.

TPI-designen delas i sin tur in, utifrån implantat–distanskopplingens position i förhållande till alveolarbenet, i ben- (bone level, BL) eller mjukvävnadsnivå (tissue level TL) och har en distansskruv av antingen titan eller keramik. Eftersom zirkonia är ett sprött material med hög elasticitetsmodul kan designen av zirkoniaimplantat och deras implantat–distanskoppling påverka belastningsöverföringen från den protetiska konstruktionen till implantatet och det omgivande benet, och därmed påverka spänningsfördelning, brotthållfasthet och i förlängningen kliniska resultat. Det finns dock begränsad forskning kring spänningsfördelning, brotthållfasthet och kliniska data för olika zirkoniabaserade designer av implantat och olika materialkombinationer.

Denna avhandling syftar till att utvärdera kliniska resultat, spänningskoncentrationer, brottshållfasthet och frakturmönster hos olika designer av dentala zirkoniabaserade implantat. I linje med avhandlingens övergripande syfte omfattar tre delstudier: Delarbete I var en systematisk översiktsstudie av kliniska studier som utvärderade kliniska resultat av olika designer av zirkoniabaserade implantat. Resultaten visade att implantat i en del (OPI) uppvisade en lovande klinisk överlevnadsgrad (95 %) efter fem år, utan några skillnader mellan olika implantatdiametrar och implantatlängder. Implantatdesignen med en syraetsad yta uppvisade högre klinisk överlevnad jämfört med implantat med andra ytor.

Delarbete II var en finit elementanalys (FEA) som utvärderade var höga spänningsnivåer och spänningskoncentrationer uppstod i olika designer av zirkoniabaserade implantatsystem. Resultaten visade att OPI- och TPI-designerna hade olika spänningsfördelningar. TPI-design uppvisade lägre spänning i det inre av själva implantatet men högre spänningsöverföring till provhållaren jämfört med OPI. I implantat-distans-komplexet uppvisade implantatutformning på mjukvävnadsnivå (TL) lägre spänning än på bennivå (BL), medan spänningarna i distansen var beroende av distansskruvens design och material. TPI med keramisk distansskruv uppvisade högre spänningsnivåer jämfört med distansskruv av titan. Delarbete III var en laboratoriestudie som utvärderade brotthållfasthet, frakturmönster och kristallfasomvandling efter dynamisk förbelastning av olika designer av zirkoniaimplantat. Resultaten visade att de utvärderade zirkoniabaserade implantatsystemen (OPI, BLTPI och TLTPI) uppvisade tillräcklig brotthållfasthet för att motivera vidare klinisk utvärdering. Designen av TPI (bennivå kontra mjukvävnadsnivå) påverkade brotthållfasthet och frakturmönster, medan distansskruvens utformning och material (keramik eller titan) hade en mindre uttalad inverkan som dessutom varierade beroende på implantatets design. Sammantaget, inom ramen för avhandlingens begränsningar, bör dentala implantat av zirkoniumdioxid inte betraktas som en enhetlig kategori. Skillnader i implantatdesign, inklusive OPI och TPI samt TL- och BL-konfigurationer och utformningen av implantat–distanskopplingen, påverkar kliniska resultat, spänningsfördelning och brotthållfasthet. Det kliniska underlaget är för närvarande bättre etablerat för OPI, som har visat lovande kliniska resultat, medan underlaget för TPI fortfarande är begränsat. Mekaniskt uppvisar TL TPI den högsta brotthållfastheten bland de utvärderade implantatdesignerna. Ytterligare långsiktiga kliniska studier behövs dock för att bekräfta dessa fynd, särskilt för TPI. 

Ort, förlag, år, upplaga, sidor
Malmö: Malmö University Press, 2026. s. 139
Serie
Malmö University Odontological Dissertations, ISSN 1650-6065, E-ISSN 2004-9307
Nationell ämneskategori
Odontologi
Identifikatorer
urn:nbn:se:mau:diva-88322 (URN)10.24834/isbn.9789178777761 (DOI)978-91-7877-775-4 (ISBN)978-91-7877-776-1 (ISBN)
Disputation
2026-10-23, Odontologiska fakulteten KL:2370 Aulan, Smedjegatan 16, Malmö, 09:15 (Engelska)
Handledare
Tillgänglig från: 2026-09-23 Skapad: 2026-09-23 Senast uppdaterad: 2026-10-27Bibliografiskt granskad

Open Access i DiVA

fulltext(1879 kB)28 nedladdningar
Filinformation
Filnamn FULLTEXT01.pdfFilstorlek 1879 kBChecksumma SHA-512
24537bfe9102752129115f4a53ef64bbabc2c0e2e932002bba3eca498dac10bc55423dd00f1e2a8a58db8d6d27c61fff6d9ac3754b3a1139417b229101a40fb1
Typ fulltextMimetyp application/pdf

Övriga länkar

Förlagets fulltextPubMedScopus

Person

Gul, AbdulazizPapia, EvaggeliaBecktor, Jonas PVult von Steyern, Per

Sök vidare i DiVA

Av författaren/redaktören
Gul, AbdulazizPapia, EvaggeliaEqtesadi, SiamakAnderson, MagnusBecktor, Jonas PVult von Steyern, Per
Av organisationen
Odontologiska fakulteten (OD)
I samma tidskrift
Clinical Oral Implants Research
Odontologi

Sök vidare utanför DiVA

GoogleGoogle Scholar
Totalt: 28 nedladdningar
Antalet nedladdningar är summan av nedladdningar för alla fulltexter. Det kan inkludera t.ex tidigare versioner som nu inte längre är tillgängliga.

doi
pubmed
urn-nbn

Altmetricpoäng

doi
pubmed
urn-nbn
Totalt: 69 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf