2020 (English)In: Dento-Maxillo-Facial Radiology, ISSN 0250-832X, E-ISSN 1476-542X, Vol. 49, no 6, article id 20190495Article in journal (Refereed) Published
Abstract [en]
OBJECTIVE: Evaluation of cone beam CT (CBCT) examination with a low-dose scanning protocol for assessment of the temporomandibular joint (TMJ).
METHODS: 34 adult patients referred for CBCT imaging of the TMJ underwent two examinations with two scanning protocols, a manufacturer-recommended protocol (default) and a low-dose protocol where the tube current was reduced to 20% of the default protocol. Three image stacks were reconstructed: default protocol, low-dose protocol, and processed (using a noise reduction algorithm) low-dose protocol. Four radiologists evaluated the images. The Sign test was used to evaluate visibility of TMJ anatomic structures and image quality. Receiver operating characteristic analyzes were performed to assess the diagnostic accuracy. κ values were used to evaluate intraobserver agreement.
RESULTS: ) averaged for the low-dose and processed protocols, according to all observers, were 0.931 and 0.941, respectively. Intraobserver agreement was good to very good.
CONCLUSION: For the CBCT unit used in this study, the low-dose CBCT protocol for TMJ examination was diagnostically comparable to the manufacturer-recommended protocol, but delivered a five times lower radiation dose. There is an urgent need to evaluate protocols for CBCT examinations of TMJ in order to optimize them for a radiation dose as low as diagnostically acceptable (the as low as diagnostically acceptable principle recommended by NCRP).
Place, publisher, year, edition, pages
British Institute of Radiology, 2020
Keywords
Cone-beam CT, Radiation dosage, TMJ